Configurations |
The system offers more than 150 scalable configurations. A configuration consists from the core unit and 1 up to 4 analytical modules (excluding the optional cobas ISE module).
The analytical modules are: cobas ISE module, cobas c 701 module, cobas c 702 module, cobas c 502 module, cobas e 801 module, cobas e 602 module. |
Detection methods |
Clinical Chemistry: Photometric
Immunochemistry: Electrochemiluminescence (ECL) |
Maximum throughput |
Up to 9800 tests/hour/configuration
- cobas ISE module: up to 1800 tests/hour
- cobas c 701 module: up to 2000 tests/hour
- cobas c 702 module: up to 2000 tests/hour
- cobas c 502 module: up to 600 tests/hour
- cobas e 801 module: up to 300 tests/hour
- cobas e 602 module: up to 170 tests/hour
|
Sample types |
Serum, Plasma, Urine, CSF, Supernatant, Whole blood (not applicable for the cobas c 701 module and the cobas c 702 module) |
Sample tubes |
13 x 75 mm, 13 x 100 mm, 16 x 75 mm, 16 x 100 mm, Hitachi standard cup 2.5 ml, cobas sample cup 2.5 ml, Hitachi microcup* 1.5 ml 11-16 x 63-102 mm (for cobas e 602 module: 13-16 mm) |
Pediatric sample tube |
Yes (Hitachi microcup*: 50 µl dead volume) |
Sample capacity |
Core unit loading capacity: 300 samples
Module Sample Buffer (MSB) capactiy: 100 samples
Core unit throughput: up to 1000 samples/hour |
Sample volume |
- cobas c 701 module: Average 5.5 µl
- cobas c 702 module: Average 5.5 µl
- cobas c 502 module: Average 7.0 µl
- cobas e 801 module: Average 19.4 µ
- cobas e 602 module: Average 26.8 µl
|
Automatic dilution |
Yes |
Reagent capacity |
Up to 280 reagent channels/configuration
- cobas ISE module: Na, K and Cl
- cobas c 701 module: 70 reagent channels
- cobas c 702 module: 70 reagent channels
- cobas c 502 module: 60 reagent channels
- cobas e 801 module: 48 reagent channels
- cobas e 602 module: 25 reagent channels
|
Reagent concept |
Ready to use |
Reagent loading |
Continuous reagent loading during operation (valid for cobas c 702 module, cobas c 502 module and cobas e 801 module) |
Sample clot detection |
Yes |
Sample interference measurement |
Serum indices (icterus, hemolysis, lipemia) measured on the clinical chemistry modules |
Remote diagnostics |
Yes |
Connectivity to smart phones and tablets |
Up to 3 systems can be connected to smart phones and tablets (iOS and Android) using cobas mobicheck application. The application allows the monitoring of the systems, alarms and reagent information. |
Dimensions (W x D x H) |
- Core unit (including STAT port): 102 x 114 x 93 cm
- cobas ISE module: 45 x 114 x 115 cm
- cobas c 701 module (including MSB): 150 x 114 x 135 cm
- cobas c 702 module (including MSB): 150 x 114 x 135 cm
- cobas c 502 module (including MSB): 150 x 114 x 130 cm
- cobas e 801 module (including MSB): 150 x 114 x 162 cm
- cobas e 602 module (including MSB): 150 x 106 x 114 cm
|
Weight |
- Core Unit: 197 kg
- cobas ISE module: 180 kg (ISE 1800), 170 kg (ISE 900)
- cobas c 701 module (including MSB): 697 kg
- cobas c 702 module (including MSB): 759 kg
- cobas c 502 module (including MSB): 535 kg
- cobas e 801 module (including MSB): 610 kg
- cobas e 602 module (including MSB): 585 kg
|
Electrical requirements |
200/208/220/230 VAC, 50/60 Hz |
Noise level |
≤ 65 dB |
Pre-Analytics connectivity |
Unidirectional, bidirectional |
Power consumption |
- Core Unit (including the control unit): 1 kVA (actual maximum: 0.4 kVA)
- cobas ISE module: 0.5 kVA (actual maximum: 0.2 kVA)
- cobas c 701 module (including MSB): 3 kVA (actual maximum: 2.3 kVA)
- cobas c 702 module (including MSB): 3 kVA (actual maximum: 2.3 kVA)
- cobas c 502 module (including MSB): 2.5 kVA (actual maximum: 1.7 kVA)
- cobas e 801 module (including MSB): 2 kVA (actual maximum: 1.5 kVA)
- cobas e 602 module (including MSB): 3 kVA (actual maximum: 1.5 kVA)
|
Heat generation |
- Core Unit (including the control unit): 3600 kJ/h (actual maximum: 1440 kJ/h)
- cobas ISE module: 1800 kJ/h (actual maximum: 720 kJ/h)
- cobas c 701 module (including MSB): 10800 kJ/h (actual maximum: 8280 kJ/h)
- cobas c 702 module (including MSB): 10800 kJ/h (actual maximum: 8280 kJ/h)
- cobas c 502 module (including MSB): 9000 kJ/h (actual maximum: 6120 kJ/h)
- cobas e 801 module (including MSB): 7200 kJ/h (actual maximum: 5400 kJ/h)
- cobas e 602 module (including MSB): 10800 kJ/h (actual maximum: 5400 kJ/h)
|